`data/fluorophores.yaml > TRITC > lifetime_ns` was `null`, so `photo/gate.py`
emitted `missing.lifetime` and lens 5 returned BLOCKED for every TRITC setting
-- no margin, no bottleneck, no difficulty grade. Filed: **2.2 ns**, free dye in
PBS, from Nanguneri et al. 2014 (Microsc Res Tech 77:510, doi
10.1002/jemt.22349).
TRITC @ 555 nm 153.0 W/cm^2 -> BLOCKED [FAIL] missing.lifetime
TRITC @ 555 nm 153.0 W/cm^2 -> PASS 10.00 perturbation.saturation
evidence: assumed advances: NO
**Two papers, and the pair is what makes it filable.** Brismar et al. 1995
(J Histochem Cytochem 43:699) measured the same quantity on a confocal laser
scanning microscope at 1.58 ns -- but with free TRITC dissolved in **ethanol**,
while putting its other three dyes in distilled water. Nanguneri measured
**2.2 ns in PBS**. This instrument's samples are aqueous, so the PBS value is
the one that transfers and the ethanol value becomes an independent cross-check
rather than a mismatch: two instruments, 572 and 575 nm, landing 1.4x apart
across a solvent change that would be expected to move it. `confidence: low`
stands -- they agree in magnitude, not to a figure.
**The two disagree about what conjugation does, which is worth more than either
number.** Brismar's IgG conjugate rises to 3.32 ns and turns double-exponential;
Nanguneri's phalloidin conjugate stays flat at 2.3 ns, with the quenching
attributed to a static dye-tryptophan complex rather than to binding. So
"conjugated" is not one condition -- and `TRITC` here is neither form:
`data/particles.yaml` records it as an explicit **stand-in** for a
rhodamine/TRITC-class bead dye whose vendor excitation and emission specs were
both measured wrong on this instrument. The registry's single name spans values
the sources measure as different, by different amounts, in different
directions, and the entry says so.
**How much the spread matters here: it does not.** G20 crosses `f = 0.1` at
`k_ex*tau = 1/9`, which puts the boundary at 5.6e4 W/cm^2 for 2.2 ns. The
measured power at the sample for the band that excites TRITC -- `Spectra >
Green`, 15.3 mW at 20x -- is 153 W/cm^2 over a 100x100 um field, 241x below it,
and the whole 1.58-3.32 ns span clamps the saturation margin at its ceiling.
Which value is filed starts to matter only near a diffraction-limited focus,
which is the tweezers regime and not this gate's. That is stated in the entry as
the reason the remaining condition mismatches are tolerable rather than
disqualifying.
`evidence: assumed`, so `advances` stays NO beside a margin of 10.00 -- the
tier's rule working as written. The falsifier is a saturation power series on
this instrument, and the entry sizes it: the departure from linear sits near
1e5 W/cm^2, so a widefield series will not reach it and the measurement needs a
deliberately tight focus.
`kb/INDEX.md` regenerated with `python -m knowledge.cli write`; 45 entries.
Suite: 1210 passed, 11 skipped.
**`filed_by` is left unfilled.** It attests that a person checked the citation
against the paper, and the draft this came from was produced by a tool that
cannot give that signature.
**Depends on the `photo/cli.py` pull request.** The before/after above cannot be
reproduced without it -- `python -m photo.cli` currently raises `TypeError` on
every invocation.
This file contains hidden or bidirectional Unicode text that may be interpreted or compiled differently than what appears below. To review, open the file in an editor that reveals hidden Unicode characters.
Learn more about bidirectional Unicode characters
Sign up for free
to join this conversation on GitHub.
Already have an account?
Sign in to comment
Add this suggestion to a batch that can be applied as a single commit.This suggestion is invalid because no changes were made to the code.Suggestions cannot be applied while the pull request is closed.Suggestions cannot be applied while viewing a subset of changes.Only one suggestion per line can be applied in a batch.Add this suggestion to a batch that can be applied as a single commit.Applying suggestions on deleted lines is not supported.You must change the existing code in this line in order to create a valid suggestion.Outdated suggestions cannot be applied.This suggestion has been applied or marked resolved.Suggestions cannot be applied from pending reviews.Suggestions cannot be applied on multi-line comments.Suggestions cannot be applied while the pull request is queued to merge.Suggestion cannot be applied right now. Please check back later.
data/fluorophores.yaml > TRITC > lifetime_nswasnull, sophoto/gate.pyemittedmissing.lifetimeand lens 5 returned BLOCKED for every TRITC setting — no margin, no bottleneck, no difficulty grade.Filed: 2.2 ns, free dye in PBS — Nanguneri et al. 2014, Microsc Res Tech 77(7):510-516, 10.1002/jemt.22349.
Two papers, and the pair is what makes it filable
Brismar et al. 1995 (10.1177/43.7.7608524) measured the same quantity on a confocal laser scanning microscope — this instrument's class — but dissolved free TRITC in ethanol while putting its other three dyes in distilled water. Samples here are aqueous, so the PBS value transfers and the ethanol value becomes an independent cross-check rather than a mismatch: two instruments, 572 and 575 nm, landing 1.4× apart across a solvent change that would be expected to move it.
confidence: lowstands. They agree in magnitude, not to a figure.The two disagree about what conjugation does, which is worth more than either number
Brismar's IgG conjugate rises to 3.32 ns and turns double-exponential. Nanguneri's phalloidin conjugate stays flat at 2.3 ns, and that paper attributes the quenching to a static dye–tryptophan complex rather than to binding.
So "conjugated" is not one condition — and
TRITChere is neither form.data/particles.yamlrecords it as an explicit stand-in for a rhodamine/TRITC-class bead dye whose vendor excitation and emission specs were both measured wrong on this instrument. One registry name spans values the sources measure as different, by different amounts, in different directions. The entry says so, and the registrynote:repeats the warning where someone reading the row will see it.How much the spread matters here: it does not
G20 crosses
f = 0.1atk_ex·τ = 1/9:The measured power at the sample for the band that excites TRITC —
Spectra > Green, 15.3 mW at 20× — is 153 W/cm² over a 100×100 µm field, 241× below the strictest of the four, and 1.5 × 10⁴ W/cm² even over a 10×10 µm field. The whole 2.1× span clamps the saturation margin at its ceiling.Which value is filed starts to matter only near a diffraction-limited focus, which is the tweezers regime and not this gate's. That is in the entry as the reason the remaining condition mismatches are tolerable rather than disqualifying — the entry states both what it buys and how coarse it is.
The tier's rule, working as written
evidence: assumed, soadvancesstays NO beside a margin of 10.00.The falsifier is a saturation power series on this instrument, and the entry sizes it: the departure from linear sits near 10⁵ W/cm², so a widefield series will not reach it and the measurement needs a deliberately tight focus.
data/light_sources.yamlalready carries the measuredpower_at_sample_mwfor both usable bands, which is what makes the falsifier something someone can actually run.Housekeeping
kb/INDEX.mdregenerated withpython -m knowledge.cli write— 45 entries. The suite caught it before I did.filed_byis left unfilled. It attests that a person checked the citation against the paper, and the draft this came from was produced by a tool that cannot give that signature. Please sign it or tell me to.🤖 Generated with Claude Code