Microresonance is an experimental non-invasive neuromuscular interface (< $100 commodity BOM) that extracts voluntary motor intent and limb posture by probing the skeletal waveguide (cortical bone) shunt-loaded by tunable viscoelastic muscle-tendon resonators rather than measuring passive electrical potentials (sEMG).
By injecting a phase-continuous 500 Hz mechanical carrier at the lumbar waist belt and performing coherent quadrature (I/Q) lock-in demodulation at a distal ankle strap (~0.9 m skeletal path) synchronized via hardware 802.11 Wi-Fi TSF (Timing Synchronization Function) clocks (< 2 µs jitter), the system achieves a resting phase noise floor of 3.01° (99.9% lock stability, 30 ms frame latency) while operating dry, through clothing (provided the emitter and receiver do not contact the same continuous piece of clothing, otherwise the mechanical carrier can propagate directly through the fabric instead of the skeletal waveguide), and remaining resilient to orthostatic postural load that saturates conventional surface EMG (20.7× baseline floor surge).
- Full 4-Page IEEE Preprint:
paper/microresonance_preprint_draft.pdf(LaTeX source) - Zenodo Archival Release (DOI): 10.5281/zenodo.24533996
- International Patent Publication: WO/2026/229867 (PCT/IB2026/054169) (priority Apr–May 2025, published Nov 5, 2026)
| Modality | Physical Observable | Skin / Coupling Requirement | Orthostatic / Postural Load Resilience | Hardware Complexity |
|---|---|---|---|---|
| Passive sEMG | Motor unit action potentials ( |
Bare skin, conductive gel/dry electrodes, low impedance |
Poor — standing postural recruitment floods the baseline ( |
Low–Medium |
| Passive MMG / Accelerometry | Low-frequency lateral fiber oscillation ( |
Dry strap | Poor — heavily corrupted by limb motion and gravity vector shifts | Low |
| Ultrasound SMG | MHz acoustic echogenicity / B-mode cross-section | Acoustic gel, rigid transducer alignment | High — measures mechanical deformation directly | High (piezo array, high-voltage pulser, FPGA/DSP) |
| Microresonance (This Work) | Coherent phase velocity & damping of a 500 Hz carrier | Dry strap / works through clothing |
High — orthogonal to electrical recruitment noise and low-frequency motion ( |
Low (ESP32-S3, MEMS ADXL355, I2S DAC exciter) |
At ~0.9 m, waist belt to distal ankle / lateral malleolus) with low loss. Think of it like playing a guitar: the emitter continuously makes the skeleton ring at a fixed carrier frequency (the vibrating string), while muscles and tendons act as tunable shunt resonators — like a finger fretting a chord on the string, altering the mechanical boundary load along the bone and shifting the wave's phase and amplitude at the distal ankle. Formally, we model the limb as a skeletal waveguide (bone) shunt-loaded by tunable viscoelastic muscle–tendon resonators (building on the biomechanical "muscle tuning" paradigm of Nigg & Wakeling, 2001):
- Letting
$Z_0 > 0$ denote the bone waveguide characteristic impedance and$Z_m(\omega; G, \eta) = R_m(\eta) + j(\omega m_{\text{eff}} - k(G)/\omega)$ the coupled muscle–tendon load impedance, the loaded transfer function and contraction phase shift follow:$$H(\omega) = \frac{H_0(\omega)}{1 + \alpha, Z_m(\omega; G, \eta)/Z_0}, \qquad \Delta\phi = \arg!\frac{H_{\text{contracted}}(\omega)}{H_{\text{relaxed}}(\omega)}$$ - Above the tens-of-Hz soft-tissue resonance (
$\omega > \omega_0$ ), increasing muscle stiffness$k(G)\uparrow$ produces a bounded phase lead ($\Delta(\arg H) = +36^\circ\text{ to }+67^\circ$ ) paired with a+11–16%transmission gain ($|H|\uparrow$ ). - Simultaneously, hip/knee joint angles alter boundary coupling between bone segments, producing sharp 3D wave polarization shifts (
$\Delta\phi_{XY}, \Delta\phi_{YZ}$ ) that classify limb posture (100%holdout accuracy,$N=1$ , zero IMU) orthogonally to voluntary isometric effort. - Because the receiver mixes the
$4000\text{ Hz}$ accelerometer stream coherently at$f_0 = 500\text{ Hz}$ against a TSF-synchronized local oscillator, low-frequency motion ($<50\text{ Hz}$ ) and stochastic sEMG firing fall outside the narrow lock-in passband.
All raw multi-modal telemetry from the reference session is committed directly in this repository:
- Raw dataset:
physio_raw_20260514_175320.jsonl(5,136frames,153.74 sduration,33.4 Hzframe rate,500.0 Hzcarrier). - Reproducible analysis script:
scripts/analyze_session.py
All four publication figures are reproducible directly from physio_raw_20260514_175320.jsonl via scripts/generate_paper_figures.py:
Six discrete biomechanical states (3 postures × relaxed/contracted) separate cleanly in 4D acoustic space (92.8% holdout accuracy, Silhouette +0.46 vs. -0.24 for sEMG); under standing orthostatic load, sEMG baseline rises 20.7× while acoustic phase preserves a 66.1° step (d = 3.57).
Independent frequency-domain SNR at the known 0.10 Hz contraction cadence (~5 s hold / ~5 s rest): Acoustic Phase achieves 33.0 / 31.3 / 28.2 dB across seated, extended, and standing regimes — outperforming co-located sEMG by +18.1 to +26.7 dB and IMU tilt by +9.8 to +12.4 dB.
Inter-axis phase differences Δφ_XY and Δφ_YZ depend on joint/waveguide geometry rather than isometric effort, achieving 100.0% IMU-free 3-posture classification (N = 1) orthogonally to voluntary contraction.
Co-located gyroscope magnitude correlates with acoustic phase velocity |dφ/dt| (r = 0.59), with 17.9× peak separation between macroscopic posture transitions (41.8 rad/s) and isometric holds (2.34 rad/s), enabling self-gating at 4 rad/s without an IMU.
| Time Window | Experimental Protocol | Acoustic Phase Behavior (500 Hz Lock-In) | Reference sEMG Behavior (Neural ADC) |
|---|---|---|---|
0 – 30 s |
Static Rest (Seated) |
Acoustic Hard-Lock. Unwrapped phase holds flat at |
Low resting baseline floor ( |
30 – 68 s |
Seated Isometric Intent Cycles (~5 s tension / relaxation) |
Clean Step Response. Voluntary isometric contractions produce repeatable, square-like phase steps (+11–16% carrier magnitude rises above |
Bursts correlate with contractions, though noisy within each hold. |
68 – 105 s |
Posture Change (Seated Leg Extension) |
Waveguide Re-Lock. At 68 s, extending the leg alters waveguide geometry; |
Baseline shifts; spike amplitude becomes less uniform. |
107 – 136 s |
Standing Up (Orthostatic Postural Load) | Preserved Intent Contrast. Despite full body-weight load on the leg, conscious isometric contractions continue to produce distinct, flat-topped phase steps (Zoom Panel 6). |
Signal Collapse. Static postural recruitment raises the sEMG baseline floor 20.7× (9.8 203.1), masking voluntary clicks in continuous background tetanus. |
138 – 153.7 s |
Seated Recalibration | Returning to the seated posture restores |
Postural tetanus subsides back to resting floor. |
-
Full Session: Neural Intent vs. Acoustic Phase: Overlays unwrapped 500 Hz acoustic phase (black, radians) against raw sEMG envelope standard deviation (red). Note the clean rectangular steps in acoustic phase throughout both seated (
30–105 s) and standing (110–136 s) regimes. -
Acoustic Transmission (Magnitude): Lock-in carrier amplitude at 500 Hz in gravitational units (
$g$ ). Above soft-tissue resonance ($\omega > \omega_0$ ), isometric stiffening increases transmission gain by+11–16%, while passive strap/respiration drift (CV = 15.6%) remains largely decoupled from resting phase (Corr(Mag, Phase) = +0.075). -
Motion Dynamics (Gyroscope): 3-axis angular velocity magnitude (
$\text{deg/s}$ ) from the co-located LSM6DSOX IMU. Confirms that phase steps during30–68 s,72–104 s, and110–135 soccur under quasi-static isometric conditions ($\sim 1^\circ/\text{s}$ ), while sharp gyroscope spikes (69 s,106 s,137 s) mark macroscopic posture transitions. -
Inter-Axis Phase Coherency (XYZ Vector Stability): Differential lock-in phase between accelerometer axes (
$\Delta\phi_{XY} = \phi_X - \phi_Y$ ,$\Delta\phi_{YZ} = \phi_Y - \phi_Z$ ). Cancels common-mode sensor translation and acts as an instantaneous geometric posture signature. -
Acoustic Reaction Speed (Phase Velocity): Time derivative
$|\text{d}\phi/\text{d}t|$ ($\text{rad/s}$ ), highlighting rapid mechanical state transitions at contraction onset/release. -
ZOOM: Standing & Final Sitting (
100 s – End): High-resolution view of the standing orthostatic stress test. Between110 sand135 s, sEMG saturates into continuous postural noise while the acoustic phase trace resolves three discrete voluntary isometric holds (114–120 s,123–130 s,130–135 s). - The Universal Bio-Acoustic Alphabet (Phase vs. Mag Space): Joint phase–magnitude state-space projection colored by sEMG intensity, revealing discrete, separable attractors corresponding to biomechanical states (relaxed seated, contracted seated, extended, standing loaded).
The hardware consists of two wireless ESP32-S3 nodes synchronized over standard 2.4 GHz Wi-Fi without any interconnecting cable between emitter and detector.
Active Acoustic Emitter Belt (microresonance_transceiver) |
Coherent Multi-Modal Receiver Strap (microresonance_receiver) |
|---|---|
![]() |
![]() |
| Node | Component | Model / Notes |
|---|---|---|
Emitter Belt (microresonance_transceiver) |
MCU | ESP32-S3 (N16R8, APLL enabled) |
| DAC | PCM5102A 32-bit I2S audio DAC (BCK/LRCK/DIN + external MCLK on GPIO 15; legacy AD9833 DDS driver also remains in code) |
|
| Audio Amplifier | ZK-502C Class-D stereo amplifier | |
| Vibration Exciter | Dayton Audio DAEX32EP-4 (4 Ω, 40 W) — much smaller exciters can also be used; the DAEX32EP-4 is oversized for low-amplitude 500 Hz probing | |
| UI (optional) | SSD1315 OLED display + 4 tactile buttons (K1–K4) |
|
Receiver Strap (microresonance_receiver) |
MCU | ESP32-S3 (FH4R2) |
| Primary Acoustic Sensor | ADXL355 3-axis 20-bit low-noise MEMS accelerometer (4000 Hz ODR via SPI + DRDY ISR) |
|
| Reference IMU | LSM6DSOX 6-axis IMU (400 kHz I2C + hardware 24-bit timestamp) |
|
| Reference sEMG | Commodity 3-contact inline gold-plated dry analog sEMG module (AliExpress, read via ESP32-S3 ADC on GPIO 1) |
|
| Experimental / Inactive | Piezo Film + ADS8681 | Driver (ads8681.h) is present in firmware from earlier tests; piezo-film experiments have not yielded useful results compared to the ADXL355. |
Achieving sub-degree acoustic phase lock across independent wireless microcontrollers required solving three hardware-level bottlenecks in the firmware:
Free-running crystal oscillators on separate MCUs drift by
-
Emitter (
microresonance_transceiver/src/main.cpp): Queries the hardware 802.11 Wi-Fi Beacon Timing Synchronization Function (esp_wifi_get_tsf_time) before each I2S DMA buffer and computes the exact analytical carrier phase for every sample$i$ at$44.1\text{ kHz}$ :$$\phi_i = 2\pi \cdot \left( \left( t_{\text{TSF}} \cdot 10^{-6} + \frac{i}{f_s} \right) f_{\text{carrier}} \bmod 1.0 \right)$$ - Driven through a 4-wire PCM5102A I2S DAC using the ESP32-S3 APLL (
use_apll = true) and external Master Clock (MCLKonGPIO 15). -
Receiver (
microresonance_receiver/src/dsp_task.cpp): Captures the rising edge of the ADXL355 hardwareDRDYpin (GPIO 13) in an IRAM ISR (drdy_isr), measures the exact ISR-to-SPI readout lagdrdyAgeUs, and back-dates the Wi-Fi TSF timestamp (tsf = tsfAfter - drdyAgeUs).
At analogRead() on the ESP32-S3 SAR ADC for sEMG measurement induces transient current spikes on the shared
-
Core 1 (
sensorTask@ Prio 5,dspTask@ Prio 4): Dedicated exclusively to deterministic$4000\text{ Hz}$ SPI bursts from the ADXL355 and non-blocking UDP dispatch. Zero ADC reads or I2C stalls occur on Core 1. -
Core 0 (
lsmTask@ Prio 3): Polls the LSM6DSOX 6-axis IMU (400 kHzI2C + hardware 24-bit timestamp registers) and samples the analog sEMG channel (GPIO 1) into a lock-free ring buffer indexed byg_active_ab.
Standard Wi-Fi MAC layers buffer outbound packets into A-MPDU aggregates (up to 16 frames), introducing
-
Wire Protocol (
microresonance_receiver/include/dsp_packet.h): Packs$120\times 3\text{-axis}$ ADXL355 samples ($30\text{ ms}$ window),$30\times 6\text{-axis}$ LSM6DSOX IMU samples + timestamps, and$120$ sEMG samples into a packed 1460-byteRawSyncPacket(0x4D58) — fitting inside a single IPv4 UDP datagram (< 1472bytes) with zero IP fragmentation. - Tags the UDP socket with
IP_TOS = 0xE0(WMMAC_VOVoice priority), instructing the Wi-Fi baseband to disable A-MPDU aggregation and transmit each$30\text{ ms}$ window immediately.
microresonance/
├── microresonance_transceiver/ # ESP32-S3 Active Acoustic Emitter Firmware (PlatformIO)
│ ├── include/ # Pinout config, TSF clock helper, AD9833 & UI state headers
│ ├── src/ # TSF-locked NCO audio engine (I2S PCM5102A), OLED UI, HTTP sync
│ └── platformio.ini # Build config for ESP32-S3-CAM (N16R8)
├── microresonance_receiver/ # ESP32-S3 Coherent Multi-Modal Receiver Firmware (PlatformIO)
│ ├── include/ # RawSyncPacket (0x4D58) wire format, DSP task & ADS8681 headers
│ ├── src/ # Dual-core ADXL355 SPI + LSM6DSOX I2C + sEMG capture & UDP streamer
│ └── platformio.ini # Build config for ESP32-S3 (FH4R2)
├── microresonance_server/ # Go Real-Time Telemetry Server, TSF Lock-In DSP & Dump Recorder
│ ├── cmd/receiverd/ # Main daemon entrypoint (UDP/TCP ingest, HTTP TSF API, Ebiten/TUI)
│ ├── internal/domain/ # RawSyncPacket decoder, TSF I/Q demodulator, JSONL physio/raw dumper
│ ├── internal/app/gui/ # Real-time dashboard (LOCK-IN, FFT, DYN, RAW, XCORR)
│ └── README.md # Server build, flags & dump recording guide
├── paper/ # 4-Page IEEE Preprint (LaTeX, 300-DPI Figures, PDF & Build Script)
│ ├── microresonance_preprint_draft.pdf # Compiled 4-page IEEE preprint
│ ├── microresonance_preprint_draft.tex # Full LaTeX source
│ ├── build_pdf.sh # Reproducible single-command PDF build script (Tectonic)
│ └── figures/ # Publication figures (300 DPI)
├── scripts/ # Python Signal Interpretation, Verification & Figure Suite
│ ├── analyze_session.py # Reproduces the 7-panel plot & regime metrics from raw JSONL
│ ├── verify_paper_claims.py # Independent 28/28 numerical audit of all paper claims vs. raw JSONL
│ ├── generate_paper_figures.py # Reproduces all 4 publication-grade 300-DPI figures
│ └── lockin_dsp_reference.py # Reference RawSyncPacket parser & TSF-coherent I/Q demodulator
├── physio_raw_20260514_175320.jsonl # Raw multi-modal experimental dataset (5,136 frames, 153.7s)
├── 1778779569118.png # Generated 7-panel visualization from Session 175320
├── IMG20260514213337.jpg # Hardware photo: Emitter belt assembly
├── IMG20260514213901.jpg # Hardware photo: Receiver / detector strap
└── requirements.txt # Python dependencies (numpy, matplotlib)
# 1. Create a virtual environment and install dependencies
python3 -m venv .venv
source .venv/bin/activate
pip install -r requirements.txt
# 2. Run the session analyzer on the raw JSONL dump (prints regime report & generates PNG)
python3 scripts/analyze_session.py \
--input physio_raw_20260514_175320.jsonl \
--output 1778779569118.png
# 3. Run the independent 28/28 numerical verification audit of all preprint claims
python3 scripts/verify_paper_claims.py
# 4. Regenerate all 4 publication-grade 300-DPI figures (paper/figures/)
python3 scripts/generate_paper_figures.py
# 5. Run the mathematical self-test of the TSF-coherent I/Q lock-in demodulator
python3 scripts/lockin_dsp_reference.py --self-test
# 6. Rebuild the 4-page IEEE preprint PDF from LaTeX source (uses Tectonic)
./paper/build_pdf.sh========================================================================================
MICRORESONANCE TELEMETRY REPORT — physio_raw_20260514_175320.jsonl
========================================================================================
Frames: 5136 | Duration: 153.74 s | Frame interval: 29.93 ms | Carrier: 500.0 Hz | Hard-Lock: 99.9%
----------------------------------------------------------------------------------------
Experimental Regime | Window (s) | Phase Range (rad) | Mean Mag (mg) | sEMG RMS/Std
----------------------------------------------------------------------------------------
1. Static Rest (Acoustic Lock) | 0.0–30.0 | 0.353 | 2.657 | 91.4
2. Seated Isometric Cycles | 30.0–68.0 | 1.093 | 3.293 | 84.0
3. Seated Leg Extension (Re-Lock) | 70.0–105.0 | 0.964 | 2.575 | 36.7
4. Standing Orthostatic Load | 110.0–136.0 | 1.910 | 2.324 | 260.8
5. Seated Recalibration | 140.0–153.7 | 0.798 | 2.106 | 50.6
----------------------------------------------------------------------------------------
Key Comparative Findings (Seated vs. Standing Orthostatic Load):
* sEMG Baseline Floor (25th pct): 9.8 (seated) -> 203.1 (standing) [20.7x postural noise surge]
* Acoustic Phase Dynamic Step: 57.8 deg (seated) vs. 103.8 deg (standing) [step contrast preserved]
* Peak Postural Transition Gyro: 159.0 deg/s (settled to 1.03 deg/s during standing holds)
========================================================================================
-
Current Validation Scope: The dataset included here represents an
$N=1$ proof-of-principle demonstrating coherent phase hard-lock and orthostatic noise immunity under controlled seated-to-standing transitions. -
Waveguide Geometry Calibration: Macroscopic changes in joint angle alter the physical path length between exciter and receiver, shifting the absolute baseline phase to a new attractor (as seen at
68 sand138 s). While inter-axis differential phase ($\Delta\phi_{XY}, \Delta\phi_{YZ}$ ) and IMU fusion cleanly segment these posture transitions, automated baseline re-zeroing across continuous dynamic motion is an active focus. -
Next Engineering Milestones:
-
Dual-Frequency MIMO Differential Probing (
f1/f2): Using beat-phase dispersion ($\Delta\phi_{\text{beat}}$ ) between two simultaneous carriers (e.g.,311 Hzand500 Hz) to cancel geometric path-length changes without relying on IMU gating. -
Multi-Subject & Multi-DoF Gesture Alphabet: Expanding state-space clustering in
$(\phi, \text{Mag}, \Delta\phi_{XY})$ space to classify fine finger and wrist isometric intents through clothing. - Single-Strap Wearable Integration: Miniaturizing the exciter and ADXL355 detector into a unified dry elastomer band for exoskeleton, prosthetics, and AR/VR human-computer interaction.
-
Dual-Frequency MIMO Differential Probing (
If you use this hardware architecture, dataset, or analysis scripts in your research, please cite:
@misc{fatkulin2026microresonance,
author = {Amir Fatkulin},
title = {Microresonance: Wireless 802.11 TSF-Coherent Active Acoustic Probing of the Skeletal Waveguide for Real-Time Isometric Motor Intent and Posture Decoupling},
year = {2026},
publisher = {Zenodo},
doi = {10.5281/zenodo.24533996},
url = {https://doi.org/10.5281/zenodo.24533996}
}non-invasive-bci human-computer-interaction biomedical-signal-processing acoustic-impedance sonomyography semg lock-in-amplifier esp32-s3 adxl355 neurotech wearable-sensors biomechanics embedded-dsp