BP-Card Framework

BP-Card is a model-card-inspired reporting framework for AI-driven cuffless blood pressure (BP) estimation, introduced in Towards Trustworthy AI-driven Cuffless Blood Pressure Monitoring (Cisnal et al., npj Digital Medicine, in press). It specifies minimum reporting requirements and separates the calibration-free and calibration-based evaluation pathways.

Worked example of the BP-Card reporting framework applied to a calibration-based cuffless blood pressure model using the Aurora-BP dataset.

Worked example of the BP-Card reporting framework applied to a calibration-based cuffless blood pressure model using the Aurora-BP dataset.

Why it exists

The paper's systematic review and meta-analysis of 86 cuffless BP studies (2017–2025) found that the field's main limitation is inconsistent, underpowered, and methodologically heterogeneous reporting rather than selective publication:

  • Pooled SBP mean error was small (1.12 mmHg, 95% CI [0.76, 1.47]) but extremely heterogeneous (I² = 98.8%).
  • Subject-wise studies showed a smaller, more stable bias (μ ≈ 0.81 mmHg) than record-wise studies (μ ≈ 1.50 mmHg) with wider dispersion — consistent with information leakage.
  • No study achieved low overall risk of bias; validation independence was the weakest domain.
  • None of the reviewed studies reported a baseline-model comparison.

BP-Card translates these findings into structured reporting requirements.

The four modules

Module Scope Rationale
0 — Clinical purpose and use context Model name and version, estimation family, calibration paradigm, model type, input data type, output variable (SBP/DBP/MAP), signal modality, sensor specification and location, intended use Operationalizes the first two taxonomy axes. The intended-use designation (research, home/ambulatory, adjunctive clinical, clinical pilot) sets the evidentiary bar for everything that follows.
1 — Dataset and cohort reporting Source (public/private), cohort size, demographics, comorbidities and medication, measurement phases, BP and ΔBP distributions, preprocessing and exclusions, distributions before/after cleaning Cohort composition and BP variability drive apparent performance. Narrow BP distributions can produce favorable error metrics that do not reflect predictive capability.
2 — Validation strategy Evaluation method, unseen-subject and external validation, reference standard and device, measurement conditions, baseline comparator Validation design determines whether performance generalizes. Subject-wise partitioning with an unseen-subject test set is the central requirement; the baseline comparison contextualizes the signal's contribution.
3A / 3B — Performance reporting Standardized metric table and recommended visualizations, stratified by BP category (3A) or calibration condition (3B), with baseline results in both A common format enables direct comparison and prevents selective metric reporting.

Core methodological requirements

  • Subject-wise validation as the minimum acceptable standard. Record-, window-, or sample-wise splits allow the same individual to appear in training and test sets, inflating apparent accuracy.
  • Calibration-free and calibration-based pathways reported separately. The calibration axis determines the appropriate validation and reporting logic.
  • Baseline-model comparison in both pathways. Calibration-based baselines use each subject's calibration BP; calibration-free baselines use population-average BP (and demographics where applicable), separating genuine signal contribution from population priors.
  • Pathway-specific evaluation conditions.
    • Calibration-free: a broad BP spectrum. IEEE 1708a-2019 requires ≥ 85 subjects with ~21 each across normotension, prehypertension, stage 1, and stage 2.
    • Calibration-based: static and dynamic testing, plus assessment of time since calibration. IEEE 1708a recommends inducing BP changes within [−30, 30] mmHg (SBP) and [−20, 20] mmHg (DBP), with at least n_total = 255 + 3·(n − 85) accuracy data points (see ieee_min_sample_size in bpcard_metrics.py).